Q
What is the limit for sterilization temperature?
The sterilization temperature must be ≤ 135 °C (or ≤ 155 °C for improved types). Exceeding this limit will cause damage to the motor.
Q
Can the battery be sterilized under high temperatures?
No, the battery is not resistant to high temperatures and cannot be sterilized. Make sure to remove the battery from the handpiece after surgery.
Q
What are the specifications for oscillations and noise level?
The device features oscillations of 11000 cpm and has a low noise emission level of ≤ 75dB during operation.
Q
How should the cranial bur be maintained after surgery?
The cranial bur should be washed immediately after the surgery with a purger to prevent blood from sticking. Ensure it is dried properly before storage.
Q
How many times can a cranial bur blade be used?
The blade is made of medical stainless steel. It is designed to be flexible rather than overly hard to avoid breakage. Due to wear and tear, we recommend using each blade for only one surgery.
Q
Are there any special handling instructions for the battery space?
Yes. After sterilization, it is crucial to ensure that the battery space inside the handpiece of the tool remains completely free of water.